A 360 degrees single-axis tilt stage for the high-voltage electron microscope.

D P Barnard1, J N Turner, J Frank

  • 1Biological Microscopy and Image Reconstruction Resource, Wadsworth Center for Laboratories and Research, University of Albany, NY 12201-0509.

Journal of Microscopy
|July 1, 1992
PubMed
Summary

A new specimen stage for high-voltage electron microscopy enables complete 180-degree tilting for advanced single-axis tomography. This innovation allows for detailed 3D reconstructions of small, cylindrical specimens, improving imaging capabilities.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Nutritional zinc deficiency mimicking acrodermatitis enteropathica in a fully breast-fed infant.

Journal of the European Academy of Dermatology and Venereology : JEADV·2006
Same author

The effect of voluntary arm abduction on balance recovery following multidirectional stance perturbations.

Experimental brain research·2006
Same author

Updates on local anaesthetics.

SADJ : journal of the South African Dental Association = tydskrif van die Suid-Afrikaanse Tandheelkundige Vereniging·2006
Same author

Why and what type of vasoconstrictors are added to local anaesthetic solutions?

SADJ : journal of the South African Dental Association = tydskrif van die Suid-Afrikaanse Tandheelkundige Vereniging·2006
Same author

Updates on local anaesthetics.

SADJ : journal of the South African Dental Association = tydskrif van die Suid-Afrikaanse Tandheelkundige Vereniging·2006
Same author

Updates on local anaesthetics.

SADJ : journal of the South African Dental Association = tydskrif van die Suid-Afrikaanse Tandheelkundige Vereniging·2006

Area of Science:

  • Microscopy
  • Materials Science
  • Biophysics

Background:

  • High-voltage electron microscopy (HVEM) is crucial for nanoscale imaging.
  • Existing specimen stages often limit tilting angles, hindering tomographic reconstruction.
  • Cylindrical specimens mounted on microneedles offer unique imaging possibilities.

Purpose of the Study:

  • To develop and characterize a novel specimen stage for HVEM.
  • To enable unrestricted rotation and a full 180-degree tilt range.
  • To assess the stage's utility for single-axis tomography and 3D reconstruction.

Main Methods:

  • Designed a spindle mechanism for coaxial mounting of microneedles/micropipettes.
  • Integrated the stage into a side-entry rod for HVEM.

Related Experiment Videos

  • Recorded 180-degree tilt series of puffball spores and processed for tomographic reconstruction.
  • Main Results:

    • The stage achieved angular accuracy within +/- 0.20 degrees.
    • Complete 180-degree tilt series allowed for high-fidelity 3D reconstructions.
    • Reconstructions showed minimal distortion down to 160 degrees tilt, with increasing distortion below that.

    Conclusions:

    • The new specimen stage significantly enhances single-axis tomography capabilities in HVEM.
    • It facilitates detailed 3D imaging of microscale cylindrical specimens.
    • The stage's performance is robust, with high-quality reconstructions achievable even with slightly reduced tilt ranges.