Related Experiment Video
Updated: Aug 23, 2026

Workflow Based on the Combination of Isotopic Tracer Experiments to Investigate Microbial Metabolism of Multiple Nutrient Sources
Published on: January 22, 2018
Differential electrolytic potentiometric determination of some amino acids in dosage forms
Sheikha M Al-Ghannam1, Abdalla Abulkibash, Abeer M Al-Olyyan
1Girls College of Science, Department of Chemistry, PO Box 838, Dammam 31113, Saudi Arabia. sm_ghannam@yahoo.com
Abstract:
The application of direct-current differential electrolytic potentiometry to the nonaqueous titration of amino acids was investigated. The basic character of amino acids in acetic acid was enhanced to permit their direct titration with perchloric acid. A pair of antimony electrodes was used as an indicating system. The shapes of the titration curves obtained were almost symmetrical with sharp peaks. The optimum current density for those titrations was found to be 1-2 microA/cm2. The procedure was applied successfully to the determination of certain amino acids in drug formulations, and the results were favorably compared statistically with those obtained by official methods.
More Related Videos
Related Concept Videos
Potentiometry: Overview
Potentiometric Titration: Overview
Potentiometry: Membrane Electrodes
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Protein-Drug Binding: Determination Methods
Indirect methods involve isolating the bound drug from its free form in biological samples such as blood, serum, or plasma. These techniques aim to measure the percentage of drugs bound to proteins. Equilibrium dialysis is a commonly used method where the free drug concentration at equilibrium is measured by separating the bound...