Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
The validity of a new noncontact tonometer and its comparison with the Goldmann tonometer
Andrew K C Lam1, Rufina Chan, Chin-hang Lam
1Department of Optometry & Radiography, The Hong Kong Polytechnic University, Kowloon, Hong Kong SAR, China. orkclam@polyu.edu.hk
Purpose:
This study compared a new noncontact tonometer (NCT), the Nidek NT-4000 (Nidek Co., Ltd., Aichi, Japan), with a pulse detection feature; an earlier model, the Nidek NT-2000, without this pulse detection facility; and Goldmann tonometry.
Methods:
Thirty-one young subjects had their intraocular pressure (IOP) in one eye randomly measured by two NCT's, followed by a Goldmann tonometer. The pulse detection mode used by the NT-4000 included synchronizing the IOP with the peak of the pulse (P-P), the middle of the pulse (P-M), or the bottom of the pulse (P-B). The order of measurements from these three modes was also random. Three consecutive readings were obtained from each mode/tonometer, and the mean was used for analysis.
Results:
Goldmann tonometry demonstrated the smallest variation from consecutive measurements (coefficient of variation, <4%), followed by the NT-4000 (coefficient of variation, <5% in various modes) and the NT-2000 (coefficient of variation, >5%). For the NT-4000, the mean IOP decreased from the highest in the P-P mode (16.1 mm Hg) to the lowest in the P-B mode (14.7 mm Hg), with the P-M mode in the middle. A significant difference was found among the IOP readings using the NT-4000 (in various modes), NT-2000, and Goldmann tonometry (repeated-measures analysis of variance, p < 0.01). More than 80% of the results from the NT-4000 were within 3 mm Hg of those from the Goldmann tonometry.
Conclusions:
The pulse detection feature from this new NCT could capture and monitor pulse waves, thus reducing the variations in the consecutive measurements. The NT-4000 also performed better than the NT-2000 when compared with the Goldmann tonometry.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Comparing Experimental Results: Student's t-Test
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...