Related Experiment Video
Updated: Jul 22, 2026

Determining Viral Disinfection Efficacy of Hot Water Laundering
Published on: June 21, 2022
Novel methods for disinfection of prion-contaminated medical devices
Guillaume Fichet1, Emmanuel Comoy, Christelle Duval
1CEA/DSV/DRM/GIDTIP, Route du Panorama, 92265 Fontenay-aux-Roses, France.
Background:
The unique resistance of prions to classic methods of decontamination, and evidence that prion diseases can be transmitted iatrogenically by medical devices pose a serious infection control challenge to health-care facilities. In view of the widespread tissue distribution of the variant Creutzfeldt-Jakob disease agent in human beings, new practicable decontamination procedures are urgently needed.
Methods:
We adapted an in-vivo method using stainless steel wires contaminated with prions to the hamster-adapted scrapie strain 263K. A new in-vitro protocol of surface contamination compatible with subsequent biochemical detection of PrP(res) (protease-resistant form of the prion protein) from the treated surface was developed to explore the mechanisms of action of methods of decontamination under test. These models were used to investigate the effectiveness of innovative physical and chemical methods of prion inactivation.
Findings:
Standard chemical decontamination methods (NaOH 1N, NaOCl 20000 ppm) and autoclaving in water at 134 degrees C reduced infectivity by >5.6 log10 lethal doses; autoclaving without immersion was somewhat less effective (4-4.5 log reduction). Three milder treatments, including a phenolic disinfectant, an alkaline cleaner, and the combination of an enzymatic cleaner and vaporised hydrogen peroxide (VHP) were also effective. VHP alone, which can be compatible with electronic components, achieved an approximately 4.5 log reduction in infectivity (equivalent to autoclaving without water immersion).
Interpretation:
New decontamination procedures are proposed to ensure the safety of medical and surgical instruments as well as surfaces that cannot withstand the currently recommended prion inactivation procedures.
More Related Videos
06:11Rapid Detection of Bacterial Pathogens Causing Lower Respiratory Tract Infections via Microfluidic-Chip-Based Loop-Mediated Isothermal Amplification
Published on: March 29, 2024
06:27Automated Hospital Room Disinfection Utilizing a Novel Aerosolized Hydrogen Peroxide Microdroplet Disbursing Technology
Published on: February 24, 2026
Related Concept Videos
Hand hygiene
Hand washing...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Chemical Agents for Microbial Control