Enhanced blood boron concentration estimation for BPA-F mediated BNCT
M Kortesniemi1, T Seppälä, I Auterinen
1HUS Helsinki Medical Imaging Center, University of Helsinki, POB 340, FIN-00029 HUS, Finland. mika.kortesniemi@hus.fi
Abstract:
The blood boron concentration regulates directly the BNCT irradiation time in which the prescribed dose to the patient is delivered. Therefore a proper estimation of the blood boron concentration for the treatment field based on the measured blood samples before irradiation is required. The bi-exponential model fit using Levenberg-Marquardt method was implemented for this purpose to provide the blood boron concentration estimates directly to the treatment data flow during the BNCT procedure. The harmonic mean bi-exponential decay half-lives of the studied patient data (n=28) were 15+/-8 and 320+/-70 min for the faster and slower half-life. The model uncertainty (n=28) was reasonably low, 0.7+/-0.1 microg/g (about 5%). The implemented algorithm provides a robust method for temporal blood boron concentration estimation for BPA-F mediated BNCT. Utilization of the infusion data improves the reliability of the estimate. The overall data flow during the treatment fulfills the practical requirements concerning the BNCT procedure.