Related Experiment Video
Updated: Aug 23, 2026

The Automated Crystallography Pipelines at the EMBL HTX Facility in Grenoble
Published on: June 5, 2021
Automation of the EMBL Hamburg protein crystallography beamline BW7B
Ehmke Pohl1, Uwe Ristau, Thomas Gehrmann
1European Molecular Biology Laboratory Hamburg Outstation, Notkestrasse 85, D-22603 Hamburg, Germany. ehmke.pohl@psi.ch
Abstract:
The EMBL Hamburg Outstation currently operates five synchrotron beamlines for protein crystallography. The strongest of these beamlines is the fixed-energy beamline BW7B which receives about half of the radiation (1.5 mrad) from a 56 pole wiggler located at the DORIS III storage ring at the German synchrotron facility DESY. Over the last years this beamline has been upgraded and equipped with a fully automated crystallographic end-station and a robotic sample changer. The current set-up allows for remote operation, controlled from the user's area, of sample mounting, centering and data collection of pre-frozen crystals mounted in Hampton-type cryovials on magnetic caps. New software and intuitive graphical user interfaces have been developed that control the complete beamline set-up. Furthermore, algorithms for automatic sample centering based on UV fluorescence are being developed and combined with strategy programs in order to further automate the collection of entire diffraction data sets.