Effective and organ doses from scanography and zonography: a comparison with periapical radiography
A Ekestubbe1, A Thilander-Klang, A Lith
1Department of Oral and Maxillofacial Radiology, Faculty of Odontology, The Sahlgrenska Academy at Göteborg University, Göteborg, Sweden. ekestubbe@odontologi.gu.se
Objectives:
To compare absorbed and effective doses from scanographic and zonographic examinations performed in the Scanora unit with intraoral periapical radiography.
Methods:
Absorbed dose measurements were made on an anthropomorphic phantom head with LiF thermoluminescent dosemeters in the regions of the pituitary gland, eye lenses, parotid glands, submandibular glands, thyroid gland and skin. Energy imparted was calculated from the measurements of air collision kerma and effective doses by using the quotient 24 mSv J-1 between energy imparted and effective dose. The upper and lower third molar region was examined with intraoral radiographs and with ramus scanograms, dental scanograms and dental zonograms. Radiation dose measurements were also performed for Scanora panoramic radiography (jaw and dental).
Results:
The effective doses for the ramus and dental scanograms were 0.01 mSv, similar or lower than for intraoral radiography. Zonography yielded the highest effective dose (0.03 mSv). Except for the skin doses, the salivary glands received the highest doses. Salivary gland doses were slightly higher from narrow beam than from intraoral radiography.
Conclusions:
Detail narrow beam radiography with the Scanora is an alternative to periapical radiography and is preferred, from a radiation dose point of view, over zonography.
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...