Related Experiment Video
Updated: Aug 7, 2026

Reservoir Condition Pore-scale Imaging of Multiple Fluid Phases Using X-ray Microtomography
Published on: February 25, 2015
Concentration of 99mTcO4-/188ReO4- by a single, compact, anion exchange cartridge
1Isotope Production Division, Pakistan Institute of Nuclear Science & Technology, P.O. Nilore, Islamabad, Pakistan. mushtaqa@pinstech.org.pk
Background:
This paper describes a simple and inexpensive post-elution anion exchange cartridge system for the concentration of 99mTcO4-/188ReO4-.
Method And Results:
The system is based on generator elution with a mixture of 0.7 M acetic acid and 0.0225 M sodium chloride instead of physiological saline to provide high specific volume solutions of 99mTc and 188Re from 99Mo-99mTc and 188W-188Re generators, respectively. The passage of the bolus through an amine-type (Accell Plus QMA Sep-Pak) anion exchange cartridge results in trapping of the microscopic levels of the pertechnetate or perrhenate. Subsequent elution of the anion exchange cartridge with 2 ml of saline then provides high specific volume solutions of pertechnetate or perrhenate.
Conclusion:
This new approach works effectively to obtain concentrated solutions of 99mTcO4- and 188ReO4- from commercially available alumina based Mo-Tc and W-Re generators, respectively.
Related Concept Videos
Expressing Solution Concentration
Concentrations may be quantitatively assessed using a wide variety of measurement units, each convenient for particular applications. Molarity (M) is a useful concentration unit for many applications in chemistry.
EDTA: Auxiliary Complexing Reagents
Extraction: Advanced Methods
Ion Exchange
Ion-Exchange Chromatography
Concentration Cells