Validation of the BodyGem hand-held calorimeter
E L Melanson1, L B Coelho, Z V Tran
1Center for Human Nutrition, University of Colorado Health Sciences Center, Denver, CO 80262, USA. Ed.melanson@uchsc.edu
The hand-held BodyGem calorimeter provides valid and reliable resting metabolic rate measurements. Adjusting for the energy cost of holding the device improves accuracy, especially for overweight and obese individuals.
Area of Science:
- Physiology
- Medical Devices
- Metabolic Research
Background:
- Accurate resting metabolic rate (RMR) measurement is crucial for metabolic research and clinical applications.
- Hand-held indirect calorimeters offer potential for portable and accessible RMR assessment.
- Validation against established methods is necessary to confirm the reliability of new devices.
Purpose of the Study:
- To evaluate the validity and reliability of a hand-held indirect calorimeter (BodyGem) for measuring RMR.
- To compare the BodyGem's RMR measurements with those obtained from a traditional metabolic cart (Sensormedics 2900).
- To assess the accuracy of the BodyGem in overweight and obese individuals compared to predictive equations.
Main Methods:
- Resting metabolic rate (RMR) was measured on two separate mornings in 41 healthy adults.
- Measurements were taken using both a metabolic cart (Sensormedics 2900) and a hand-held indirect calorimeter (BodyGem).
- The energy cost of holding the hand-held device was quantified and applied as a correction factor.
Main Results:
- No significant trial-to-trial differences were observed for either device.
- The BodyGem initially reported significantly higher RMR values than the Sensormedics 2900.
- After adjusting for the energy cost of holding the device, the differences between the two systems became non-significant.
- The BodyGem demonstrated higher accuracy than the Harris-Benedict equations for estimating RMR in overweight and obese individuals.
Conclusions:
- The hand-held BodyGem indirect calorimeter is a valid and reliable tool for RMR measurement.
- The higher readings initially observed with the BodyGem are attributable to the energy expenditure required to hold the device.
- The BodyGem offers a more accurate assessment of RMR in overweight and obese populations compared to traditional predictive formulas.
More Related Videos
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
09:33Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024
Related Concept Videos
Constant Pressure Calorimetry
Constant Volume Calorimetry
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Calorimetry