Abstract:
We describe an unusual complication during flotation of a pulmonary artery catheter through a preexisting percutaneous introducer sheath. A malfunctioning pulmonary artery catheter, which was placed through an introducer sheath in the right internal jugular vein, was removed. Attempts at repositioning a second pulmonary artery catheter met with resistance, and we were unable to either advance or withdraw it. Chest radiograph showed a bent introducer sheath going from the right internal jugular vein into the right subclavian vein and a pulmonary artery catheter loop. Under continuous fluoroscopy, the introducer sheath and the pulmonary artery catheter were withdrawn as one unit, which resulted in relaxation of the acute angulation in the introducer sheath and allowed the pulmonary artery catheter to unfold, thus facilitating their complete extraction. We conclude that complications may occur during placement of a pulmonary artery catheter through a well positioned introducer sheath and that fluoroscopy is a valuable tool for safe management of such a complication.
Insights
A pulmonary artery catheter complication occurred when a second catheter could not be advanced or withdrawn. Fluoroscopy guided removal of the catheter and introducer sheath as one unit, resolving the issue.
Area of Science:
- Cardiology
- Medical Devices
- Interventional Procedures
Background:
- Pulmonary artery catheters are crucial for hemodynamic monitoring.
- Percutaneous introducer sheaths facilitate catheter placement.
- Complications during insertion can impede patient care.
Observation:
- A malfunctioning pulmonary artery catheter was removed via a right internal jugular vein introducer sheath.
- A second pulmonary artery catheter became lodged, resistant to advancement or withdrawal.
- Chest radiograph revealed a bent introducer sheath and a looped pulmonary artery catheter.
Findings:
- The introducer sheath and pulmonary artery catheter were successfully withdrawn as a single unit under fluoroscopy.
- Withdrawal allowed the introducer sheath to straighten, releasing the catheter loop.
- Complete extraction was facilitated by managing the angulated sheath and looped catheter.
Implications:
- Complications can arise even with well-positioned introducer sheaths during pulmonary artery catheter placement.
- Continuous fluoroscopy is essential for managing such procedural difficulties.
- This technique ensures safe extraction, minimizing patient risk.