Related Experiment Video
Updated: Aug 22, 2026

Use of MRI-ultrasound Fusion to Achieve Targeted Prostate Biopsy
Published on: April 9, 2019
System for prostate brachytherapy and biopsy in a standard 1.5 T MRI scanner
Robert C Susil1, Kevin Camphausen, Peter Choyke
1Department of Biomedical Engineering, Johns Hopkins University School of Medicine, Baltimore, MD, USA.
Abstract:
A technique for transperineal high-dose-rate (HDR) prostate brachytherapy and needle biopsy in a standard 1.5 T MRI scanner is demonstrated. In each of eight procedures (in four patients with intermediate to high risk localized prostate cancer), four MRI-guided transperineal prostate biopsies were obtained followed by placement of 14-15 hollow transperineal catheters for HDR brachytherapy. Mean needle-placement accuracy was 2.1 mm, 95% of needle-placement errors were less than 4.0 mm, and the maximum needle-placement error was 4.4 mm. In addition to guiding the placement of biopsy needles and brachytherapy catheters, MR images were also used for brachytherapy treatment planning and optimization. Because 1.5 T MR images are directly acquired during the interventional procedure, dependence on deformable registration is reduced and online image quality is maximized.
Related Concept Videos
Imaging Studies IV: Magnetic Resonance Imaging
Magnetic Resonance Imaging