Temperature requirements for thermal modulation in comprehensive two-dimensional gas chromatography
Richard B Gaines1, Glenn S Frysinger
1U.S. Coast Guard Academy, Department of Science, 27 Mohegan Ave., New London, CT 06320-8101, USA. gaines@dcseq.uscga.edu
Abstract:
Temperature requirements for trapping and release of compounds in a cryogenic gas loop-type GC x GC modulator were determined. Maximum trapping temperatures on the uncoated, deactivated modulator capillary were determined for compounds from C4 (bp -0.5 degrees C) to C40 (bp 522 degrees C). The liquid-nitrogen cooled gas flow rate was reduced from a high of 15.5 to 1.5 SLPM over the range to achieve the required trapping temperature. Excessive cold jet flow rates caused irreversible trapping and peak tailing for semi-volatile compounds above C26. Alternate cold jet coolants were investigated. An ice water-cooled jet was able to trap compounds with boiling points from C18 (bp 316 degrees C) to C40 and a room temperature air-cooled jet was able to trap compounds from C20 (bp 344 degrees C) to C40. The hot jet produced launch temperatures approximately 40 degrees C hotter than the elution temperature with heating time constants of 8 to 27 ms. Modulated compound peaks were symmetrical with half-height peak widths of 43 to 56 ms for compounds with little second column retention, and 70 to 75 ms for compounds with more second column retention. The liquid nitrogen-cooled loop modulator with gas flow programming was used to produce a GC x GC chromatogram for a crude oil that contained compounds from C7 to C47.
More Related Videos
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
10:14Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
Related Concept Videos
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...