Related Experiment Video
Updated: Aug 22, 2026

Athymic Rat Model for Evaluation of Engineered Anterior Cruciate Ligament Grafts
Published on: March 26, 2015
Early tibial tray failure of a Duracon knee with retrieval analysis
Ta-Feng Ho1, Ruey-Yug Tsai, Pei-Yuan Lee
1Department of Orthopaedics Surgery, Show Chwan Memorial Hospital, Changhua City, Changhua County, Taiwan.
Abstract:
We report a case of early tibial tray fracture of a Duracon knee prosthesis. Aside from the clinical, radiographic, and gross analysis of the failed prosthesis, we also performed analysis on the retrieved polyethylene component and the fractured tibial baseplate. In the analyses, we noted significant femoral component malalignment, uneven tray cementation, and inherent metallurgical weakness. It appears that the high compressive load on the medial tray resulted in bending fatigue failure. To avoid this complication, it is important to restore the normal alignment of the knee joint and use a polyethylene insert of higher conformity, at least 6 mm thick.