Direct assessment of profilometric roughness variability from typical implant surface types
Sean S Kohles1, Melissa B Clark, Christopher A Brown
1Kohles Bioengineering, Portland, Oregon 97214-5135, USA. ssk@kohlesbioengineering.com
The International Journal of Oral & Maxillofacial Implants
|September 7, 2004
Summary
Surface characterization instruments for dental implants show varied accuracy and precision. Instrument choice significantly impacts roughness measurements, potentially affecting osseointegration understanding.
Area of Science:
- Biomaterials Science
- Surface Engineering
- Dental Implantology
Background:
- Surface roughness is critical for dental implant osseointegration.
- Current methods for quantifying implant surface roughness lack standardization.
- Variability in measurement protocols can lead to inconsistent data.
Purpose of the Study:
- To evaluate the accuracy and precision of common roughness characterization instruments used in the dental implant industry.
- To compare the performance of interferometers and stylus profilometers for measuring average roughness (Ra).
- To assess the influence of different surface treatments on roughness measurements.
Main Methods:
- Four types of dental implant surfaces (machined, acid-etched, hydroxyapatite-coated, titanium plasma-sprayed) were analyzed.
- Average roughness (Ra) was measured using an interferometer and a stylus profilometer.
- Scanning electron microscopy (SEM) was used for cross-sectional topographic analysis.
- Analysis of variance (ANOVA) was employed to determine the significance of instrument type, surface treatment, and measurement parameters.
Main Results:
- Both surface treatment and instrument type significantly influenced Ra measurements (P < .0001).
- The interferometer demonstrated higher accuracy in predicting SEM-based roughness compared to the stylus profilometer.
- The stylus profilometer exhibited superior aggregate precision (lower coefficient of variation) over all measurements.
Conclusions:
- Accuracy and precision of surface roughness quantification depend heavily on the chosen characterization technique.
- Instrument variability in roughness measurement may complicate the establishment of correlations between implant surface topography and osseointegration.
- Standardized protocols for roughness measurement are needed to ensure reliable data in dental implant research.