Friction in hip-joint prostheses and its influence on the fixation of the artificial head
R Schäfer1, U Soltész, P F Bernard
1Fraunhofer-Institut für Werkstoffmechanik, Wöhlerstrasse 11, D-79108 Freiburg, Germany. RS@IWM.FHG.DE
Abstract:
The head of an implanted hip joint endoprosthesis is exposed to torques, which are transferred during gait due to the friction between the head and the cup prosthesis. In prostheses with ceramic ball heads, which are widely used now, and in which the head is fixed onto the stem by conical clamping, these torques could possibly affect the connection. In this study, torques transferred from the cup to the head are compared to the torques which are required to loosen the head from the metallic spigot. The results show that for the investigated head and taper types and sizes, under normal conditions the connection is safe with respect to undesired rotation. However, it is shown that for polluted sliding surfaces the fixation strength could possibly be exceeded.