Quantitative assessment of truncal FDG-PET examination with postinjection transmission scan--comparison with
Tatsuro Tsuchida1, Katsuya Sugimoto, Sadahiko Nishizawa
1Department of Radiology Faculty of Medicine, University of Fukui, Japan. tsucchy@fmsrsa.fukui-med.ac.jp
Unlabelled:
The purpose of this study was to assess the quantitative accuracy of truncal FDG PET with a postinjection transmission scan.
Methods:
Ten subjects with lung cancer were recruited for this study. Prior to the emission scan, a transmission scan was performed for 10 min. All subjects received 370 MBq of intravenous administration of FDG prior to a 60-min emission scan. Immediately following the emission scan, a postinjection transmission scan was performed. Emission data from 40 to 60 min postinjection were reconstructed with either pre- or postinjection transmission data and converted to a standardized uptake value (SUV) image. On each SUV image, 5 regions of interest were placed and regions of interest on the SUV image with a postinjection transmission scan (SUVpost) were plotted against those with preinjection transmission (SUVpre), and a regression line was generated. Using the slope and Y-intercept of the regression line, percent error of estimation of the SUV was calculated based on the following equation: % error = ISUVpre - SUVpostI x 100/SUVpre.
Results:
In the low SUV area (SUV = 1), the averaged percent error was 9.4 +/- 12.0% (mean +/- SD), whereas in the high SUV area (SUV = 10), the averaged percent error was 2.8 +/- 3.1%. The least percent error was 1.8 +/- 1.8% (SUV = 3.8) in this study.
Conclusion:
In the study on truncal FDG PET with postinjection transmisson scan, the quantitative accuracy was preserved and the method is clinically available.
More Related Videos
15:10A Dual Tracer PET-MRI Protocol for the Quantitative Measure of Regional Brain Energy Substrates Uptake in the Rat
Published on: December 28, 2013
10:04Analysis of 18FDG PET/CT Imaging as a Tool for Studying Mycobacterium tuberculosis Infection and Treatment in Non-human Primates
Published on: September 5, 2017
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
![Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography (FDG-PET/CT)](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F3777.jpg&w=3840&q=50)