Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Radiation doses in interventional radiology procedures: the RAD-IR Study. Part III: Dosimetric performance of the
Stephen Balter1, Beth A Schueler, Donald L Miller
1Department of Medicine, Lenox Hill Hospital, New York, NY, USA.
Purpose:
To present the physics data supporting the validity of the clinical dose data from the RAD-IR study and to document the performance of dosimetry-components of these systems over time.
Materials And Methods:
Sites at seven academic medical centers in the United States prospectively contributed data for each of 12 fluoroscopic units. All units were compatible with International Electrotechnical Commission (IEC) standard 60601-2-43. Comprehensive evaluations and periodic consistency checks were performed to verify the performance of each unit's dosimeter. Comprehensive evaluations compared system performance against calibrated ionization chambers under nine combinations of operating conditions. Consistency checks provided more frequent dosimetry data, with use of each unit's built-in dosimetry equipment and a standard water phantom.
Results:
During the 3-year study, data were collected for 48 comprehensive evaluations and 581 consistency checks. For the comprehensive evaluations, the mean (95% confidence interval range) ratio of system to external measurements was 1.03 (1.00-1.05) for fluoroscopy and 0.93 (0.90-0.96) for acquisition. The expected ratio was 0.93 for both. For consistency checks, the values were 1.00 (0.98-1.02) for fluoroscopy and 1.00 (0.98-1.02) for acquisition. Each system was compared across time to its own mean value. Overall uncertainty was estimated by adding the standard deviations of the comprehensive and consistency measurements in quadrature. The authors estimate that the overall error in clinical cumulative dose measurements reported in RAD-IR is 24%.
Conclusion:
Dosimetric accuracy was well within the tolerances established by IEC standard 60601-2-43. The clinical dose data reported in the RAD-IR study are valid.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies VII: Vascular Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography