(HP)Ge Measurement of spectra for diagnostic X-ray beams
M S Nogueira1, H C Mota, L L Campos
1Centro de Desenvolvimento da Tecnologia Nuclear, R. Mário Werneck, S/N, 30123-970, Belo Horizonte, MG, Brazil. mnogue@cdtn.br
Abstract:
The X-ray spectra applied in diagnostic radiology to the energy range between 50 and 125 kVp, were experimentally determined using a high-purity germanium detector, (HP)Ge, coupled to a multichannel analyser. A simple stripping procedure was implemented, according to that described by Seelentag and Panzer, to correct the pulse height distribution and then to determine the photon spectra. The measurements were performed for those conditions where the X-ray beam is transmitted through Pb filters 0.2-0.7 mm thick. It was necessary to estimate the effective dose in the same way that it is done for radiology diagnosis room shielding projects. The spectra were also determined to the X-ray qualities recommended by the IEC for primary diagnostic X-ray and for X-ray beams attenuated by aluminium filters 2.5-42.5 mm thick. The spectra obtained were compared with data derived from the literature presenting good agreement.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Gas Chromatography: Types of Detectors-II