Performance evaluation of a specific IgE assay developed for the ADVIA centaur immunoassay system
Anita Birgit Petersen1, Pernille Gudmann, Pernille Milvang-Grønager
1In Vitro Diagnostics Business Unit, ALK-Abelló, Denmark. AIP@dk.alk.abello.com
Objective:
To develop and evaluate a liquid phase immunoassay for accurate determination of allergen-specific IgE (sIgE) as a useful tool in the diagnosis of allergy patients.
Design And Methods:
A fully automated, quantitative sIgE assay was developed for the ADVIA Centaur technology platform using a unique calibration method based on a recombinant reference allergen. Compared to most other IgE-assays, the assay employs a reverse sandwich architecture using monoclonal mouse anti-human IgE antibody covalently bound to paramagnetic particles in the solid phase and capturing the sample IgE. Bound sIgE reacts with liquid biotin-labeled allergen, which is detected as chemiluminescence using acridiniumester-labeled streptavidin.
Results:
The ADVIA Centaur sIgE assay (Centaur assay) has exclusive reactivity to human IgE and performs with excellent linearity in the assay range 0.35-100 kU/L and high precision (imprecision within-run <2.6%, between-run <4.9%, and total imprecision <7.1%). The analytical sensitivity is <0.10 kU/L. Using Pharmacia CAP system FEIA (CAP) as a comparative method, positive/negative concordance was 94% at 0.35 kU/L cut-off, and the Centaur assay has a sensitivity of 90% and a specificity of 98%. Validation of the assay in a general population sample (The Copenhagen allergy study) revealed that sIgE was highly associated with a clinical diagnosis of inhalation allergy.
Conclusions:
The Centaur assay is an allergen-specific assay for measurement of IgE without interference from other types of immunoglobulins or nonspecific IgE. The assay performs with a linear reaction, high assay range, and good reproducibility. The assay correlates well with the CAP system and is in agreement with clinical diagnosis.
More Related Videos
08:26Development and Validation of an Ultrasensitive Single Molecule Array Digital Enzyme-linked Immunosorbent Assay for Human Interferon-α
Published on: June 14, 2018
07:52Detection of Antibodies That Neutralize the Cellular Uptake of Enzyme Replacement Therapies with a Cell-based Assay
Published on: September 10, 2018
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Automated Microbial Diagnostics