The impact of hemolysis on Ortho-Clinical Diagnostic's ECi and Roche's elecsys immunoassay systems
J A Snyder1, M W Rogers, M S King
1Department of Pathology and Laboratory Medicine, University of North Carolina, Chapel Hill, NC, USA.
Background:
Hemolysis is regularly encountered in clinical specimens and often interferes with a variety of laboratory test methods. Although not widely recognized, immunoassays based on nonisotopic detection systems can also be affected by hemolysis. For this reason, we investigated the effect of differing amounts of hemolysis across a range of values for several immunoassays on the Ortho-Clinical Diagnostics ECi and Roche Elecsys platforms.
Methods:
Hemolysate was prepared from whole blood and spiked at varying concentrations into pooled patient serum samples for different analytes.
Results:
Out of the 21 analytes tested, six (28.6%) exhibited significant increases or decreases in measured concentrations with increasing amounts of hemolysis.
Conclusions:
Although immunoassays are generally thought to be impervious to hemolysis interference, hemolysis can interfere in immunoassay testing platforms. For these reasons, we recommend that laboratories conduct hemolysis interference studies for all laboratory test protocols.
More Related Videos
09:32Procedure and Key Optimization Strategies for an Automated Capillary Electrophoretic-based Immunoassay Method
Published on: September 10, 2017
06:50A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...
Automated Microbial Diagnostics