Influence of blood collection techniques on platelet function.

H Mani1, K Kirchmayr, C Kläffling

  • 1Department of Internal Medicine, University Hospital Frankfurt/Main, Germany. helenmani@gmx.de

Platelets
|September 17, 2004
PubMed
Summary

Butterfly cannulae do not affect platelet function tests. This study found no significant differences in platelet aggregation or PFA-100 results when using butterfly cannulae versus standard needles for blood collection.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

It's time to improve antithrombotic therapy! A global perspective for antithrombotic stewardship.

European journal of internal medicine·2026
Same author

New recommendations on cerebral venous and dural sinus thrombosis from the German consensus-based (S2k) guideline.

Neurological research and practice·2024
Same author

Diagnostics and Therapy of Venous Thrombosis and Pulmonary Embolism. The revised AWMF S2k Guideline

VASA. Zeitschrift fur Gefasskrankheiten·2023
Same author

Adherence to Statin Therapy Drives Survival of Patients with Symptomatic Peripheral Artery Disease.

European cardiology·2020
Same author

Hamostaseologie·2019
Same author

[Modern treatment of deep vein thrombosis and pulmonary embolism].

Der Internist·2019

Area of Science:

  • Hematology
  • Clinical Pathology
  • Biomedical Engineering

Background:

  • Platelet function is crucial for hemostasis.
  • Previous recommendations advised against butterfly cannulae for platelet function testing due to potential platelet activation.
  • The impact of different blood collection systems on platelet function assays requires clarification.

Purpose of the Study:

  • To evaluate the influence of butterfly cannulae versus standard needle systems on platelet function test results.
  • To determine if butterfly cannulae affect platelet activation or aggregation.
  • To assess the reliability of platelet function assays with different blood collection devices.

Main Methods:

  • Blood samples were collected simultaneously from 25 healthy volunteers using a 21-gauge needle and a 21-gauge butterfly cannula system.

Related Experiment Videos

  • Platelet function was analyzed using the PFA-100 (platelet function analyzer) with COL/EPI and COL/ADP cartridges.
  • Platelet aggregation was measured using the Behring Coagulation Timer (BCT) and PAP-4 optical aggregometer with ADP, collagen, and arachidonic acid stimulation.
  • Main Results:

    • No significant differences were observed in closure times on the PFA-100 between butterfly cannulae and standard needles.
    • Optical aggregometry results, including maximum aggregation response, showed no significant variations between the two collection systems.
    • Platelet function assays (aggregometry and PFA-100) were not significantly affected by the type of blood collection system used.

    Conclusions:

    • Butterfly cannulae are suitable for blood collection in platelet function investigations.
    • The use of butterfly cannulae does not appear to activate platelets or interfere with standard platelet function assays.
    • Current recommendations against butterfly cannulae for platelet function testing may be re-evaluated based on these findings.