Tube and column agglutination technology for autocontrol testing
J E Courtney1, J L Vincent, A J Indrikovs
1University of Texas Medical Branch, Galveston, TX 77555, USA.
Immunohematology
|September 18, 2004
Summary
Positive autocontrols in blood testing are common with column agglutination technology, particularly the gel test. Further investigation is needed to understand the clinical significance of these positive results.
Area of Science:
- Transfusion Medicine
- Immunology
- Clinical Pathology
Background:
- Positive autocontrol results in blood group serology can indicate alloimmunization or autoantibodies.
- Column agglutination technology (CAT) offers increased sensitivity compared to traditional methods.
- Variability in positive autocontrol rates between different CAT systems necessitates further study.
Purpose of the Study:
- To investigate the incidence and significance of positive autocontrols using the ID Micro Typing System (gel) and Gamma ReACT (ReACT) column agglutination technologies.
- To compare the frequency of positive autocontrols between gel, ReACT, and conventional tube methods.
- To evaluate the clinical relevance of positive autocontrols, especially when direct antiglobulin test (DAT) is negative.
Main Methods:
- Prospective study of 1021 patient and 95 donor samples.
- Autocontrol testing performed using ID Micro Typing System (gel), Gamma ReACT, and low-ionic-strength solution (LISS) tube methods.
- Direct antiglobulin test (DAT) performed via tube method, with elution studies as needed.
Main Results:
- A 7.74% incidence of positive autocontrols was observed in patient samples.
- The gel test showed the highest frequency of positive autocontrols (91.13% of positive cases), followed by ReACT (26.58%) and tube method (34.18%).
- Of 79 positive autocontrols, 44 had a negative DAT, and only one sample with a positive DAT had a clinically significant antibody (anti-D).
Conclusions:
- Column agglutination techniques, especially the gel method, exhibit higher sensitivity for detecting positive autocontrols than the conventional tube method.
- Significant differences exist in the frequency of positive autocontrols between ReACT and gel systems.
- Further investigation into the clinical significance of positive autocontrols in CAT, particularly when the tube method is also positive, is recommended.