Related Experiment Video
Updated: Aug 11, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
The characterization of a promising new optical source for use with a radiotherapy treatment simulator
J Theal1, K Leszczynski, L Reed
1Northeastern Ontario Regional Cancer Centre, Sudbury, Ontario P3E 5J1, Canada.
Abstract:
We report on an experimental examination of a new powerful optical source with potential application in radiotherapy treatment simulation. The illuminance, effective source position, and source size have been measured for this new source and, for comparison, for conventional sources on a simulator and a linear accelerator. This new source is considerably brighter than those in common use in radiotherapy. Its effective optical source size is larger than that on the linear accelerator tested but comparable with that on a conventional simulator.