Related Experiment Video
Updated: Aug 5, 2026

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Applicability of CORVUS pencil beam model and scatter dose model for intensity modulated neutron therapy
1Barbara Ann Karmanos Cancer Institute, Gershenson Center, Harper Hospital, Wayne State University, 48201 MI, USA.
Abstract:
Intensity modulated neutron radiotherapy (IMNRT) is currently being investigated as a mechanism to improve dose conformality in neutron radiotherapy, thereby minimizing normal tissue toxicity. This study investigates the applicability of two different dose calculation algorithms for IMNRT, a commercial system which utilizes a finite size pencil beam (FSPB) model, and an in-house planning system which uses a differential scatter air ratio (DSAR) method. Calculated dose distributions were compared with measured profiles for validation purposes. The beam-profiles matched to within +/-3% in the central region of the field. The 80-20% penumbra width as measured using an ionization chamber varied as 0.6 cm and 1.0 cm for 3 x 3 and 10 x 10 cm2 profile at a depth of 2.5 cm. The FSPB model fitted the data to a penumbra width of 0.1 cm for both 3 x 3 and 10 x 10 cm2 profiles. These results indicate that the commercial system needs further investigation. However, the in-house planning system has been validated for small irregular fields for IMNRT to an accuracy of +/-5%. Absolute dose measurements agreed with the calculated doses to within +/-3%.

