Related Experiment Video
Updated: Aug 21, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Preliminary in-hospital experience with a fully automatic external cardioverter-defibrillator
André Moreira Bento1, Luiz Francisco Cardoso, Sérgio Timerman
1Valvular Heart Disease Unit, Instituto do Coração (InCor), University of São Paulo Medical School, Av Enéas de Carvalho Aguiar, 44, São Paulo, SP 05403-000, Brazil. andrebento@cardiol.br
Background:
Ventricular fibrillation (VF) and ventricular tachycardia (VT) are frequently present as initial rhythms during in-hospital cardiac arrest. Although ample evidence exists to support the need for rapid defibrillation, the response to in-hospital cardiac arrest remains without major advances in recent years. The delay between the arrhythmic event and intervention is still a challenge for clinical practice.
Objective:
To analyze the performance and safety of in-hospital use of a programmable, fully automatic external cardioverter-defibrillator (AECD).
Methods:
We conducted a prospective study at the Emergency Department of a university hospital. A total of 55 patients considered to be at risk of sustained VT/VF were included. Patients underwent monitoring of their cardiac rhythm by the AECD. Upon detection of a ventricular tachyarrhythmia, the AECD was programmed to automatically deliver shock therapy.
Results:
We recorded 19 episodes of VT/VF in 3 patients. The median time between the beginning of the arrhythmia and the first defibrillation was 33.4 s (21-65 s). One episode of spontaneous reversion of VT was documented 20 s after its origin and shock therapy was aborted. The defibrillation success was 94.4% (17/18) for the first shock and 100% (1/1) for the second shock. No case of inappropriate shock discharge was registered during the study period.
Conclusion:
The AECD has the feasibility to combine long-term monitoring with automatic defibrillation safely and effectively. It presents the possibility of providing rapid identification of, and response to, in-hospital ventricular tachyarrhythmias.