Related Experiment Video
Updated: Aug 21, 2026

An Anaerobic Biosensor Assay for the Detection of Mercury and Cadmium
Published on: December 17, 2018
Analysis of inorganic mercury species associated with airborne particulate matter/aerosols: method development
Xinbin Feng1, Julia Y Lu, D Conrad Grègoire
1Department of Chemistry, Biology and Chemical Engineering, Ryerson University, 350 Victoria Street, Toronto, ON M5B 2 K3, Canada.
Abstract:
This paper describes a method for speciation of Hg associated with airborne particulate matter. This method uses a mini-sampler for sample collection and analysis, thermal desorption for separating Hg species, and inductively coupled plasma mass spectrometry (ICP-MS) for identification and quantification of Hg. Coal fly ash spiked with different Hg compounds (e.g. Hg0, HgCl2, HgO, and HgS) was used for qualitative calibration. A standard reference material with a certified value for Hg concentration was used to evaluate the method. When the temperature of the furnace was programmed at a linear rate of increase of 50 degrees min(-1), different Hg compounds could clearly be separated. Three airborne particulate matter samples were collected in parallel in Toronto, ON, Canada and analyzed using this method. Reproducible results were obtained and Hg0, HgCl2, HgO, and HgS species from these samples were detected.
More Related Videos
Related Concept Videos
Atomic Absorption Spectroscopy: Atomization Methods
Sample Preparation for Analysis: Advanced Techniques
Acid digestion with strong acids is commonly used to dissolve inorganic materials that are insoluble (do not dissolve) in water. This method can be useful for...
Gravimetry: Overview
In precipitation gravimetry, the analyte is converted into a precipitate and weighed. For example, the silver content in a sample can be estimated by precipitating and...