Related Experiment Video
Updated: Aug 21, 2026

Nanomechanics of Drug-target Interactions and Antibacterial Resistance Detection
Published on: October 25, 2013
Impact of nano- and mesoscale particles on the performance of microcantilever-based sensors
Lawrence A Bottomley1, Mark A Poggi, Shanxiang Shen
1School of Chemistry & Biochemistry, Georgia Institute of Technology, Atlanta, GA 30332-0400, USA. Lawrence.bottomley@chemistry.gatech.edu
Abstract:
Microcantilever-based sensors comprise an emerging class of chemomechanical sensors. The crucial challenge for every new and promising sensing platform lies in its performance in complex mixtures. Since most biofluids are rich in particulates, we assessed the impact of particles in the liquid stream on the performance of microcantilever sensors operated in both deflection and resonance modes. For both detection modes, sensor response depends on the particle diameter, concentration, and velocity as well as the composition of a thin-film coating. The presence of particles in the fluid stream produce substantial scattering of the laser beam used to measure cantilever deflection. Thus, prior removal of particulate matter from biofluids is required for optimal performance of microcantilever-based biosensors.