Inaccuracies in MTS assays: major distorting effects of medium, serum albumin, and fatty acids

Kuang Tzu Huang1, Yen Hao Chen, Ameae M Walker

  • 1University of California, Riverside, CA 92521-0121, USA.

Biotechniques
|October 9, 2004
PubMed

Insights

Soluble formazan assays for cell counting are unreliable due to media interference and serum albumin effects. A new protocol provides accurate cell number assessment by overcoming these limitations.

Area of Science:

  • Biotechnology
  • Cell Biology
  • Biochemistry

Background:

  • Soluble formazan assays are standard for cell number determination.
  • Previous studies have noted discrepancies between assay readings and actual cell counts.

Purpose of the Study:

  • To investigate the causes of inaccurate readings in soluble formazan assays.
  • To develop an improved protocol for reliable cell number assessment.

Main Methods:

  • Investigated the influence of culture media components on formazan production.
  • Assessed the impact of serum albumin and fatty acids on assay absorbance.
  • Developed and validated a new protocol for soluble formazan assays.

Main Results:

  • Culture media components amplify formazan production, confounding results.
  • Serum albumin significantly reduces assay absorbance, obscuring cell number changes.
  • Fatty acid concentration modulates the albumin-mediated reduction in absorbance.

Conclusions:

  • Standard soluble formazan assays are prone to significant inaccuracies.
  • A novel protocol effectively corrects for media and serum interferences.
  • The improved assay provides consistent and representative cell number measurements.