Improved method of in vivo respiratory-gated micro-CT imaging
Erin B Walters1, Kunal Panda, James A Bankson
1Department of Imaging Physics, Unit 56, The University of Texas M D Anderson Cancer Center, Houston, TX 77030, USA.
Abstract:
The presence of motion artifacts is a typical problem in thoracic imaging. However, synchronizing the respiratory cycle with computed tomography (CT) image acquisition can reduce these artifacts. We currently employ a method of in vivo respiratory-gated micro-CT imaging for small laboratory animals (mice). This procedure involves the use of a ventilator that controls the respiratory cycle of the animal and provides a digital output signal that is used to trigger data acquisition. After inspection of the default respiratory trigger timing, we hypothesized that image quality could be improved by moving the data-acquisition window to a portion of the cycle with less respiratory motion. For this reason, we developed a simple delay circuit to adjust the timing of the ventilator signal that initiates micro-CT data acquisition. This delay circuit decreases motion artifacts and substantially improves image quality.