Related Experiment Video
Updated: Aug 21, 2026

Determination of Cardiac Output in a Porcine Model for Ex Vivo Pulmonary Perfusion
Published on: June 28, 2024
Non-invasive measurement of cardiac output in patients with chronic heart failure
Stephen J Leslie1, Sinéad McKee, David E Newby
1Department of Medical Sciences, The University of Edinburgh, Western General Hospital, Edinburgh EH4 2XU, Scotland, UK. s.j.leslie@ed.ac.uk
Objectives:
The measurement of cardiac output by thoracic bioimpedance has been previously assessed in several studies. However, there continues to be disagreement as to whether this technique is sufficiently accurate for use in clinical practice or research. The current study aimed to compare thoracic bioimpedance (COTB) with thermodilution (COTD) in patients with stable chronic heart failure.
Methods And Results:
A total of 282 paired measurements of cardiac output from 11 patients were analysed. There was good correlation between COTB and COTD (r=0.76, P<0.0001). However, Bland-Altman analysis revealed an average difference between values of 0.3 (2.2) l/min (P=0.02), suggesting a small average bias but marked variability in results. There was no significant correlation when results were expressed as percentage change from baseline and a significant average difference between values of 10.1 (30.1)%. There was no difference in between-day repeatability between thermodilution and thoracic bioimpedance [-0.2 (1.2) versus 0.1 (1.0) l/min, P=0.7].
Conclusions:
This study demonstrates a correlation between the techniques but shows a poor level of agreement. The method of COTB underestimated cardiac output compared with COTD, and this difference appeared greater with higher cardiac outputs. Agreement was worse when results were expressed as change from baseline. The present study does not support the use of thoracic bioimpedance in its current form as an alternative to thermodilution in stable patients with chronic heart failure.
Related Concept Videos
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Heart Failure VII: Nursing Interventions
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...