Accuracy and precision of quantitative energy-dispersive x-ray spectrometry in the environmental scanning electron
Robert A Carlton1, Charles E Lyman, James E Roberts
1Rhone-Poulenc Rorer, Collegeville, Pennsylvania, USA. robert.a.carlton@gsk.com
Abstract:
The accuracy and precision of quantitative energy-dispersive x-ray spectrometry in the environmental scanning electron microscope have been estimated using a series of copper / gold alloys of known composition. The mean values (five to six replicate experiments) had relative errors within +/- 5%, and most were within +/- 3.5%. All relative standard deviations were < 5% and most were < 3%. Since the standard specimens were large (approximately 500 microm) in diameter, electron scattering in the 2 torr of water vapor above the specimen did not affect the results. This level of accuracy and precision was possible only by using a novel specimen surface charge neutralization scheme.