Related Experiment Video
Updated: Aug 21, 2026

The Supraclavicular Fossa Ultrasound View for Central Venous Catheter Placement and Catheter Change Over Guidewire
Published on: December 23, 2014
ECG-guided central venous catheter positioning: does it detect the pericardial reflection rather than the right
W Schummer1, C Schummer, A Müller
1Friedrich-Schiller-University of Jena, Department of Anaesthesia and Intensive Care Medicine, Jena, Germany. cwsm.schummer@gmx.de
Background And Objective:
Although electrocardiography (ECG) guidance of central venous catheters (CVCs) is traditionally thought to detect the entrance into the right atrium (RA), there is little evidence in the literature to confirm this. We previously observed a high incidence of left-sided CVCs abutting the wall of the superior vena cava (SVC), even when the catheters were advanced past the point of increased P-wave amplitude. Our hypothesis was that this ECG amplitude signal is actually detecting the pericardial reflection rather than the RA. The goal of the study was to position catheter tips under ECG guidance outside the RA.
Methods:
One-hundred central venous triple-lumen catheters inserted either via the right or the left internal jugular veins, respectively, were analysed in cardiac surgical patients. The position of the catheter tip was ascertained by ECG. METHOD A: A Seldinger guide-wire in the distal lumen served as exploring electrode, the respective insertion depth was recorded. METHOD B: The middle lumen (port opening 2.5 cm from the catheter tip, thus the catheter was advanced more towards the atrium) filled with a saline 10% fluid column served as the exploring electrode, and the insertion depth was recorded again. Descriptive data are given as mean+/-standard deviation.
Results:
On average, the catheters were advanced by the expected 2+/-0.3 cm using Method B beyond the initial insertion by Method A. All 100 CVCs were finally correctly positioned in the SVC and confirmed by transoesophageal echocardiography. When chest radiography was performed after surgery not a single catheter abutted the lateral wall of the SVC.
Conclusion:
Since both methods detected the same structure, and catheters placed by Method B did not result in intra-atrial CVC tip position, the first increase in P-wave amplitude does correspond to a structure in the SVC, most likely the pericardial reflection.
More Related Videos
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...