Pulse dye densitometry using indigo carmine is useful for cardiac output measurement, but not for circulating blood

Y Fujita1, T Yamamoto, M Fuse

  • 1Kawasaki Medical School, Departments of Anesthesiology and ICM, Japan. fujitay@med.kawasaki-m.ac.jp

Insights

A new indigo carmine densitometer accurately measures cardiac output but overestimates blood volume. This method shows promise for cardiovascular monitoring, though further research is needed for precise volume assessment.

Area of Science:

  • Cardiovascular Physiology
  • Medical Instrumentation
  • Diagnostic Techniques

Background:

  • Cardiac output and circulating blood volume are critical hemodynamic parameters.
  • Accurate measurement is essential for patient management and research.
  • Existing methods have limitations, necessitating novel approaches.

Purpose of the Study:

  • To evaluate the validity of a novel pulse dye densitometer using indigo carmine.
  • To assess its accuracy in measuring cardiac output and circulating blood volume.
  • To compare its performance against established techniques.

Main Methods:

  • Cardiac output and blood volume were measured in dogs using the indigo carmine densitometer.
  • Measurements were taken during varying volemic states (normovolaemia, hypovolaemia, hypervolaemia).
  • Validation involved comparison with thermodilution (cardiac output) and 51Cr-labelled red blood cells (blood volume).

Main Results:

  • The indigo carmine densitometer showed good agreement with thermodilution for cardiac output (r = 0.885).
  • Circulating blood volume measurements by the densitometer were higher and showed weaker agreement (r = 0.587) compared to the 51Cr method.
  • Indigo carmine demonstrated significant passage into the extravascular space, indicated by a sieving coefficient of 0.34 during haemofiltration.

Conclusions:

  • Indigo carmine densitometry is a reliable method for determining cardiac output.
  • The method tends to overestimate circulating blood volume.
  • Extravascular distribution of indigo carmine likely accounts for the overestimation of blood volume.
Abstract