Related Experiment Videos
Integrated protein microchip assay with dual fluorescent- and MALDI read-out
D Finnskog1, A Ressine, T Laurell
1Department of Electrical Measurements, Lund Institute of Technology, Lund University, P.O. Box 118, S-221 00 Lund, Sweden.
Abstract:
A pore chip protein array (PCPA) concept based on a dual readout configuration, fluorescence imaging, and MALDI-TOF MS has been developed. Highly packed, (>4000 spots/cm2), antibody arrays were dispensed on the porous chip by using a piezo-electric microdispenser. Sandwich assay was made after blocking by addition of a secondary antibody either IgG-FITC-labeled or anti-Ang II. The antigen in the first system was a large protein (IgG), and in the other system, a FITC marked peptide Angiotensin II (Ang II) was used. Ang II antibodies showed specificity for Ang II, while the Ang I antibodies showed binding properties for Ang I, II, and Renin. Fluorescence and MALDI TOF MS read-out was made for IgG and Ang II. A major advantage of the dual read-out PCPA approach is that both affinity binding and mass identity are derived. Detection limits for Ang II on the chip is as low as 500 zmol (Ang II).