Related Experiment Video
Updated: Aug 21, 2026

Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Discrimination of osteoporotic patients with quantitative ultrasound using imaging or non-imaging device
Géraldine Falgarone1, Raphaël Porcher, Agnès Duché
1Centre d'Evaluation des Maladies Osseuses, Institut de Rhumatologie, Hôpital Cochin, Université René Descartes, 27, rue du Faubourg St Jacques, 75014 Paris, France.
Objectives:
Quantitative ultrasound (QUS) has emerged as a new tool in the assessment of fracture risk. The aim of this study was to compare the clinical utility of QUS parameters measured using imaging and non-imaging devices in the discrimination of osteoporotic patients.
Methods:
QUS (Broadband Ultrasound Attenuation, BUA dB/MHz, and Speed of Sound, SOS m/s) were measured and then statistical analyses were performed.
Results:
The 106 women included were 65 +/- 8 years aged. Using DXA, T score was < or = -2.5 at either lumbar spine or hip in 59% of patients, and 25% had osteoporotic fractures. QUS results were different among devices, and these differences were highly dependent on the measured value. There was a similar effect of age and duration of menopause for all parameters. To obtain 90% of sensitivity for the diagnosis of osteoporosis, the thresholds were 50.80 and 71.70 dB/MHz for BUA and 1544.80 and 1551.50 m/s for SOS, using imaging and non-imaging devices, respectively. Belonging to the highest tertile of QUS had a negative predictive value for osteoporosis ranging from 59% to 65%. In the lowest tertile of QUS, the proportion of osteoporotic women was between 73% and 80%. All QUS parameters, except BUA measured with the non-imaging device, were able to discriminate post-menopausal women with fractures after adjustment for age and hip BMD.
Conclusions:
Our data suggest that an imaging system improves the utility of BUA measurement, but not SOS, for post-menopausal osteoporosis assessment.
More Related Videos
07:12Semiautomated Longitudinal Microcomputed Tomography-based Quantitative Structural Analysis of a Nude Rat Osteoporosis-related Vertebral Fracture Model
Published on: September 28, 2017
08:52Software-Assisted Quantitative Measurement of Osteoarthritic Subchondral Bone Thickness
Published on: March 18, 2022
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
X-ray Imaging