Related Experiment Video
Updated: Aug 15, 2026

Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Comparison of four demand oxygen delivery systems at rest and during exercise for chronic obstructive pulmonary
C Fuhrman1, C Chouaid, R Herigault
1Service de Pneumologie, Centre Hospitalier Intercommunal de Créteil, Créteil, France.
Objectives:
The aim of this study was to assess the performance of four demand oxygen delivery systems (DODS) in improving oxygenation and effort tolerance, at rest and during exercise, in chronic obstructive pulmonary disease (COPD) patients.
Materials And Methods:
Thirteen COPD patients were prospectively included. Four DODS (Oxiclip, Versatile, Venture and Impulse) were compared with continuous-flow oxygen (CFO). Nine of these patients performed 6-min walking tests on room air and on 3 l/min oxygen by DODS and CFO; Oxygen saturation, walking distance and the Borg dyspnea score were recorded.
Results:
With all four DODS devices arterial oxygenation was improved with lower oxygen flow rates than with CFO. Oxygen economy was best with Impulse, but at a cost of less satisfactory oxygenation. Exercise desaturation was similar with CFO, Oxiclip, Venture, and Impulse but significantly higher with Versatile (P < 0.05). Borg dyspnea scores were similar with CFO, Oxiclip, Venture, and Versatile but worse with Impulse (P < 0.05). There was no significant difference in walking distances.
Conclusions:
All four DODS improved oxygen saturation and saved oxygen. However, performance was better with the two devices (Oxiclip and Venture) that deliver a bolus of oxygen at inspiration onset.
More Related Videos
Related Concept Videos
Acute Respiratory Failure-IV
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Chronic Obstructive Pulmonary Disease-V: Management
Smoking Cessation
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: