Continuous positive airway pressure delivered with a "helmet": effects on carbon dioxide rebreathing
Paolo Taccone1, Dean Hess, Pietro Caironi
1Department of Anesthesia and Critical Care, Massachusetts General Hospital, Harvard Medical School, Boston, MA, USA.
Objective:
The "helmet" has been used as a novel interface to deliver noninvasive ventilation without applying direct pressure on the face. However, due to its large volume, the helmet may predispose to CO2 rebreathing. We hypothesized that breathing with the helmet is similar to breathing in a semiclosed environment, and therefore the PCO2 inside the helmet is primarily a function of the subject's CO2 production and the flow of fresh gas through the helmet.
Design:
Human volunteer study.
Setting:
Laboratory in a university teaching hospital.
Subjects:
Eight healthy volunteers.
Interventions:
We delivered continuous positive airway pressure (CPAP) with the helmet under a variety of ventilatory conditions in a lung model and in volunteers.
Measurements And Main Results:
Gas flow and CO2 concentration at the airway were measured continuously. End-tidal PCO2, CO2 production, and ventilatory variables were subsequently computed. We found that a) when CPAP was delivered with a ventilator, the inspired CO2 of the volunteers was high (12.4 +/- 3.2 torr [1.7 +/- 0.4 kPa]); b) when CPAP was delivered with a continuous high flow system, inspired CO2 of the volunteers was low (2.5 +/- 1.2 torr [0.3 +/- 0.2 kPa]); and c) the inspired CO2 calculated mathematically for a semiclosed system model of CO2 rebreathing was highly correlated with the values measured in a lung model (r = .97, slope = 0.92, intercept = -1.17, p < .001) and in the volunteers (r = .94, slope = 0.96, intercept = 0.90, p < .001).
Conclusions:
a) The helmet predisposes to CO2 rebreathing and should not be used to deliver CPAP with a ventilator; b) continuous high flow minimizes CO2 rebreathing during CPAP with the helmet; and c) minute ventilation and Pco2 should be monitored during CPAP with the helmet.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Cardiopulmonary Resuscitation II: ACLS Airway Management
Acute Respiratory Failure-III
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...