Related Experiment Video
Updated: Aug 21, 2026

A Bedside, Single Burr Hole Approach to Multimodality Monitoring in Severe Brain Injury
Published on: March 26, 2019
Accuracy and stability of temperature probes for intracranial application
Beat Alessandri1, Bernd M Hoelper, Robert Behr
1Department of Neurosurgery, Institute of Neurosurgical Pathophysiology, Johannes Gutenberg-University of Mainz, Langenbeckstrasse 1, 55101 Mainz, Germany. beat.alessandri@uni-mainz.de
Abstract:
Intracranial temperature measurement may play a pivotal role for prognosis and treatment of neurological and neurosurgical patients. For reliable clinical application, accurate temperature readings are therefore necessary. We present an independent in vitro study investigating the accuracy and stability of three temperature probes. Eight Neurovent-P Temp (RN), eight Licox temperature sensors (LT) and eight Neurotrend sensors (NT) were placed into a water bath. The temperature was increased in 3 degrees C increments from 30 to 42 degrees C before (accuracy test day 0) and after (accuracy test day 5) a long-term stability test run at 37 +/- 0.2 degrees C. The accuracy tests revealed deviations of <0.25, <0.2 and >0.4 degrees C for the RN, NT and LT probes, respectively, when compared to the reference measurement by a precision Pt100 temperature measuring instrument. All sensor types showed stable readings over the course of 120 h. The high variability of LT probes was due to a malfunctioning Licox monitor. Excluding these values reduced deviation below 0.21 degrees C the standard deviation at each temperature step was below +/-0.08 (RN, NT) and +/-0.12 (LT), laying within the range provided by the manufacturer (RN, NT: +/-0.1; LT: +/-0.2). In general, RN, NT and LT temperature measurement is reliable, but malfunctioning parts may lead to false interpretation of temperature readings. Therefore, validation of temperature probes to a reference temperature prior to clinical use is recommended.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...