Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
A SEM evaluation of debris removal from endodontic files after cleaning and steam sterilization procedures
D A Van Eldik1, P S Zilm, A H Rogers
1Dental School, The University of Adelaide.
Background:
In recent times, it has been proposed to classify endodontic files as single-use items due to a perceived inability to adequately clean the instruments. The purpose of the present study was to quantify the surface debris on files removed from the manufacturer's packaging, and after cleaning using an ultrasonic bath or a thermal disinfector.
Methods:
Stainless steel and rotary nickel-titanium files were examined after removal from the manufacturer's packaging, after instrumentation in broth-contaminated human teeth, and after various cleaning procedures. The cleaning procedures consisted of either a thermal disinfector cycle, ultrasonication with the files placed in a perforated container or ultrasonication with the files loosely placed in a beaker. The presence of manufacturing debris and biological debris was evaluated using scanning electron microscopy and quantified using image analysis software.
Results:
The effectiveness of cleaning was not affected by variation in the size or taper of the files when an effective cleaning procedure was used. Cleaning the files in a thermal disinfector or by ultrasonication within a container did not consistently achieve complete removal of biological debris. Placing the files loosely in the ultrasonic bath achieved the most effective cleaning, an average of 98.33 per cent of the file surface area was freed of any biological debris.
Conclusions:
A conventional cleaning method is capable of effectively removing biological debris from endodontic files. The efficacy of ultrasonic cleaning was impaired when the files were placed within a perforated container.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...