Related Experiment Video
Updated: Aug 21, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Emission probabilities of K X- and gamma-rays following 51Cr and 67Ga decay
1Department of Primary Education, Faculty of Erzincan Education, Atatürk University, 24030 Erzincan, Turkey. pasayalcin@hotmail.com
Abstract:
Emission probabilities of K X- and gamma-rays following the decay of 51Cr and 67Ga were precisely measured with a calibrated Si(Li) detector. The photopeak efficiency of the Si(Li) semiconductor detector was determined experimentally for about 4-400 keV energy range by using standard sources. Results of this study have been compared with the theoretical values and other available experimental results in the literature. Good agreement was observed between our results, and other available experimental results and theoretically calculated values.
Related Concept Videos
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay:
Radioactive Decay and Radiometric Dating
Biological Effects of Radiation
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Nuclear Transmutation