Related Experiment Video
Updated: Aug 21, 2026

Ex Vivo Porcine Experimental Model for Studying and Teaching Lung Mechanics
Published on: April 19, 2024
Comparison of three manual ventilation devices using an intubated mannequin
S G Hussey1, C A Ryan, B P Murphy
1Erinville Hospital, Cork, Ireland.
Objective:
To compare three devices for manual neonatal ventilation.
Design:
Participants performed a two minute period of ventilation using a self inflating device, an anaesthesia bag with attached manometer, and a Neopuff device. An intubated neonatal mannequin, approximating a 1 kg infant with functional lungs, was used for the study. Target ventilation variables included a rate of 40 breaths per minute, peak inspiratory pressure (PIP) of 20 cm H2O, and positive end expiratory pressure (PEEP) of 4 cm H2O. The circuit was attached to a laptop computer for data recording.
Results:
Thirty five participants were enrolled, including consultant neonatologists, paediatricians, and anaesthetists, paediatric and anaesthetic registrars, and neonatal nurses. The maximum PIP recorded using the self inflating bag, anaesthetic bag, and Neopuff device were 75.9, 35.5, and 22.4 cm H2O respectively. There were significant differences between the devices for mean PIP (30.7, 18.1, and 20.1 cm H2O), mean PEEP (0.2, 2.8, and 4.4 cm H2O), mean airway pressure (7.6, 8.5, and 10.9 cm H2O), % total breaths < or = 21 cm H2O PIP (39%, 92%, and 98%), and % total breaths > or = 30 cm H2O PIP (45%, 0, and 0). There was no difference between doctors and allied health professionals for the variables examined.
Conclusion:
The anaesthetic bag with manometer and Neopuff device both facilitate accurate and reproducible manual ventilation. Self inflating devices without modifications are not as consistent by comparison and should incorporate a manometer and a PEEP device, particularly when used for resuscitation of very low birthweight infants.
More Related Videos
04:30Endotracheal Intubation Using a Flexible Intubation Endoscope as a Standardized Model for Safe Airway Management in Swine
Published on: August 25, 2022
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation I: Indication and Settings
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...