Related Experiment Video
Updated: Aug 21, 2026

Preparation and Pathogen Inactivation of Double Dose Buffy Coat Platelet Products using the INTERCEPT Blood System
Published on: December 7, 2012
Paired comparison of Gambro Trima Accel versus Baxter Amicus single-needle plateletpheresis
Edwin A Burgstaler1, Jeffrey L Winters, Alvaro A Pineda
1Division of Transfusion Medicine, Mayo Clinic, Rochester, Minnesota 55905, USA.
Background:
Gambro BCT recently introduced the Trima Accel Version 5 (TR) plateletpheresis machine. Platelet (PLT) yields, collection efficiencies (CEs), numbers of white blood cells (WBCs), and processing times of the TR versus the Amicus (Version 2.51) single-needle (AM) procedures were evaluated by use of a prospective paired comparison.
Study Design And Methods:
Target yields of 3.0 x 10(11) to 6.8 x 10(11) PLTs in up to 100 minutes of processing time were used. To detect a difference of 1.0 x 10(11) PLTs with a power of 80 percent, 26 paired comparisons were needed.
Results:
The mean amount of whole blood processed was significantly higher for TR than for AM (3795 vs. 3520 mL). The TR and AM were equivalent in regard to mean preprocedure PLT count (259 x 10(9) vs. 251 x 10(9)/L), PLT yields (6.7 x 10(11) vs. 6.5 x 10(11)), split rate (65% vs. 65%), processing time (73 vs. 78 min), and collection rate (0.090 x 10(11) vs. 0.084 x 10(11) PLTs/min). The TR had a significantly lower CE than the AM (76% vs. 86%). All of the products (after splitting) had fewer than 5 x 10(6) WBCs.
Conclusions:
The Trima Accel machine processed significantly more whole blood with equivalent PLT yields, processing time, and number of PLTs per minute compared to the Amicus single-needle procedure, but had a significantly lower CE.