Related Experiment Video
Updated: Aug 18, 2026

Veno-Venous Extracorporeal Membrane Oxygenation in a Mouse
Published on: October 24, 2018
Management of continuous venous gas emboli during extracorporeal life support utilizing the Kolobow gas trap
Sherry C Faulkner1, Charles E Johnson, Juan Tucker
1Arkansas Children's Hospital, Little Rock, AR 72202-3591, USA.
Abstract:
Extracorporeal life support (ECLS) with a roller pump system uses a closed cardiopulmonary bypass (CPB) circuit not equipped with a venous reservoir. Hence, gas emboli cannot escape the ECLS circuit, predisposing to clot formation, membrane failure and potential gas embolism. Rarely, some patients may develop a continuous release of gas into the venous circulation from multiple sources. Two pediatric ECLS cases are presented with continuous venous gas embolism. A 'gas trap' was devised by creating a column of fluid erected vertically on the venous line. This allowed gas to rise within the column, separating it from the ECLS circuit, thus, preventing gas from lodging in the membrane.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Venous Thrombosis III: Interprofessional Care
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care