Verification of implantable cardioverter defibrillator (ICD) interventions by nonlinear analysis of heart rate
Andrzej Przybylski1, Rafał Baranowski, Jan Jacek Zebrowski
1Institute of Cardiology, Warsaw, Poland. przybylski@eko.net.pl
Aims:
Conventional ICD algorithms yield approximately 10-30% of spurious interventions. Our aim was to check whether nonlinear dynamics methods might be useful in the verification of ICD interventions.
Methods And Results:
We extracted 190 consecutive RR files (approximately 2000-9000 RR intervals long) from the ICDs of 70 patients (36 with coronary artery disease, 8 with hypertrophic cardiomyopathy, 19 with dilated cardiomyopathy and 7 with other diseases). The 3D phase space trajectories in delay coordinates, window pattern entropy, and algorithmic complexity of the RR intervals were examined within a 50 beat sliding window. Data were not filtered for arrhythmia and artefacts. Of the 83 recordings with appropriate interventions 79 were correctly recognised in a blind test. Two interventions were not identified in patients with fast atrial fibrillation and two in cases of complex and frequent forms of arrhythmia. There were nine spurious interventions. In all except one case (atrial fibrillation with a fast ventricular response) the analysis by nonlinear methods showed that the intervention was not necessary. All of the 98 control recordings were correctly identified in the blind test.
Conclusions:
The results show that nonlinear dynamics methods may be used to supplement the existing ICD detection algorithms to enhance the detection success rate.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...