Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Four-dimensional CT scans for treatment planning in stereotactic radiotherapy for stage I lung cancer
René W M Underberg1, Frank J Lagerwaard, Johan P Cuijpers
1Radiation Oncology, VU University Medical Center, Amsterdam, The Netherlands.
Purpose:
Hypofractionated stereotactic radiotherapy (SRT) for Stage I non-small-cell lung cancer requires that meticulous attention be paid toward ensuring optimal target definition. Two computed tomography (CT) scan techniques for defining internal target volumes (ITV) were evaluated.
Methods And Materials:
Ten consecutive patients treated with SRT underwent six "standard" rapid multislice CT scans to generate an ITV(6 CT) and one four-dimensional CT (4DCT) scan that generated volumetric datasets for 10 phases of the respiratory cycle, all of which were used to generate an ITV(4DCT). Geometric and dosimetric analyses were performed for (1) PTV(4DCT), derived from the ITV(4DCT) with the addition of a 3-mm margin; (2) PTV(6 CT), derived from the ITV(6 CT) with the addition of a 3-mm margin; and (3) 6 PTV(10 mm), derived from each separate GTV(6 CT), to which a three-dimensional margin of 10 mm was added.
Results:
The ITV(4DCT) was not significantly different from the ITV(6 CT) in 8 patients, but was considerably larger in 2 patients whose tumors exhibited the greatest mobility. On average, the ITV(6 CT) missed on average 22% of the volume encompassing both ITVs, in contrast to a corresponding mean value of only 8.3% for ITV(4DCT). Plans based on PTV(4DCT) resulted in coverage of the PTV(6 CT) by the 80% isodose in all patients. However, plans based on use of PTV(6 CT) led to a mean PTV(4DCT) coverage of only 92.5%, with a minimum of 77.7% and 77.5% for the two most mobile tumors. PTVs derived from a single multislice CT expanded with a margin of 10 mm were on average twice the size of PTVs derived using the other methods, but still led to an underdosing in the two most mobile tumors.
Conclusions:
Individualized ITVs can improve target definition for SRT of Stage I non-small-cell lung cancer, and use of only a single CT scan with a 10-mm margin is inappropriate. A single 4D scan generates comparable or larger ITVs than are generated using six unmonitored rapid CT scans, a finding related to the ability to account for all respiration-correlated mobility.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography