Calculating Heart Rate Variability from ECG Data from Youth with Cerebral Palsy During Active Video Game Sessions
Published on: June 5, 2019
Reliability of three commercially available heart rate variability instruments using short-term (5-min) recordings
G R H Sandercock1, P Bromley, D A Brodie
1Research Centre for Health Studies, Buckinghamshire Chilterns University College, Chalfont St Giles, Buckinghamshire HP8 4AD, UK. gsande01@bcuc.ac.uk
Aim:
To assess the reliability of heart rate variability (HRV) measures made by three commercially available analysers in healthy subjects.
Methods:
Twenty-nine volunteers (20 males, mean age 35 +/- 13 years and nine females, mean age 29 +/- 11 years) underwent repeated HRV measures under three conditions: lying supine, standing, lying supine with controlled breathing. HRV was measured simultaneously by three instruments. Reliability was assessed statistically by calculating coefficient of variation (CV), intraclass correlation coefficient (ICC) and limits of agreement (LoA).
Results:
A wide range of values were found for CV (1-235%) and ICC (R = 0.16-0.99) dependent on the HRV measure assessed and the position in which the measurement was made. For the most part the analysers gave similar values in each condition. The values for CV and ICC were high but within the range reported in the literature. Values for LoA were also high and showed a wide range of values.
Conclusions:
The similarity in measures between systems indicates that biological variation and experimental error play a major role in determining the repeatability of HRV measurements. It is therefore recommended that population-specific reliability coefficients should be published where possible and that authors should take into account the reliability of measures when making sample size calculations.
Related Concept Videos
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring pulse