Operator dependent variability in quantitative analysis of myocardial perfusion images
L Akesson1, A Svensson, L Edenbrandt
1Department of Clinical Physiology, Malmö University Hospital, SE-205 02 Malmö, Sweden.
Clinical Physiology and Functional Imaging
|November 4, 2004
Summary
Experienced operators showed significant variability in quantifying myocardial perfusion defects, potentially impacting clinical decisions. However, left ventricular ejection fraction (LVEF) quantification demonstrated low inter-operator variability.
Area of Science:
- Nuclear Cardiology
- Medical Imaging Analysis
Background:
- Myocardial perfusion imaging (MPI) is crucial for diagnosing coronary artery disease.
- Accurate quantification of perfusion defects and left ventricular function is essential for patient management.
Purpose of the Study:
- To assess inter-operator variability in quantifying myocardial perfusion defects and left ventricular ejection fraction (LVEF) using two common software programs.
- To determine the impact of operator variability on clinical interpretation of MPI studies.
Main Methods:
- Eight experienced technologists processed MPI studies from 13 patients using Cedars Emory (CEqual) and Cedars-Sinai quantitative gated single-photon emission tomography programs.
- Manual steps included defining left ventricle axes and selecting apex/base for CEqual analysis.
- Quantification of perfusion defect size, reversibility, and LVEF was performed.
Main Results:
- Significant inter-operator variability was observed in quantifying reversible apical perfusion defects across all vascular territories.
- A >10% variability in reversibility was noted in at least one territory for 9 of 13 patients.
- Inter-operator variability for LVEF quantification was consistently low (<5%) across all patients.
Conclusions:
- High variability in quantifying reversible perfusion defects may lead to misinterpretation and false clinical conclusions.
- Low variability in LVEF quantification suggests its reliability across different operators.
- Standardization of manual steps in MPI analysis is needed to improve consistency.