Related Experiment Video
Updated: Aug 21, 2026

A Straightforward Method for Glucosinolate Extraction and Analysis with High-pressure Liquid Chromatography (HPLC)
Published on: March 15, 2017
High-performance liquid chromatographic method for simultaneous determination of baicalein and baicalein
1School of Pharmacy, Faculty of Medicine, The Chinese University of Hong Kong, Shatin, New Territories, Hong Kong, SAR, PRC.
Abstract:
A simple HPLC method with ultraviolet detection was developed for the simultaneous determination of baicalein and its major metabolite, baicalein 7-glucuronide (baicalin) in rat plasma. Following solid phase extraction with HLB cartridge, the analytes were separated using a gradient mobile phase consisting of methanol-acetonitrile-phosphate buffer. The flow-rate was set at 1ml/min and the eluent was detected at 320nm. The method is linear over the studied range of 1-10 and 0.05-1microg/ml for baicalein and baicalin, respectively. The intra-day and inter-day variations of the analysis were less than 6.38%, with relative error ranging from -3.6 to 6%. The limit of quantification in plasma was 0.05microg/ml for both baicalein and baicalin. The developed method proves to be an improved and more reliable method because of its capability for determining baicalein and baicalin simultaneously in a single chromatographic run.