Related Experiment Video
Updated: Aug 21, 2026

Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
Oxidized zirconium femoral components reduce polyethylene wear in a knee wear simulator
Kace A Ezzet1, Juan C Hermida, Clifford W Colwell
1Scripps Clinic, Division of Orthopaedic Surgery, La Jolla, CA 92037, USA. kezzet@scrippsclinic.com
Abstract:
Polyethylene wear remains a major problem that can jeopardize the long-term durability of prostheses used in total knee arthroplasty. Oxidized zirconium is a material that combines the strength of a metal with the wear properties of a ceramic. This study evaluated the wear rates of polyethylene inserts against oxidized zirconium femoral components. Three oxidized zirconium femoral components and three Co-Cr femoral components of identical geometry used in total knee arthroplasty were articulated against standard tibial components with modular tibial inserts made of noncross-linked ultra-high molecular weight polyethylene. Gravimetric and volumetric polyethylene wear rates were calculated after 5 million gait cycles on an AMTI knee wear simulator. Oxidized zirconium reduced polyethylene wear by 42% compared with Co-Cr alloy. This study shows that oxidized zirconium can reduce polyethylene wear substantially when used for fixed-bearing total knee arthroplasty.