Fast GC analysis with a 50 microm ID column: theory, practical aspects, and application to a highly complex sample
Luigi Mondello1, Peter Quinto Tranchida, Alessandro Casilli
1Dipartimento Farmaco-chimico, Facoltà di Farmacia, Università di Messina, viale Annunziata, 98168 Messina, Italy. lmondello@pharma.unime.it
Abstract:
This research focuses on the minimization of GC analysis times through the use of a 5 m x 0.05 mm ID x 0.05 microm (film thickness) column. Experimental minimum plate height (Hmin) and optimum linear velocity values were derived from standard compound applications, under various analytical conditions, and then related to classical chromatographic theory. Deviations from the latter are measured and discussed. Practical aspects linked to the use of such capillaries, such as column sample capacity and detector acquisition rates, are also considered. Furthermore, a fast, and what can be considered a very fast method, were applied to the separation of a fuel sample. Coefficients of variation of elution times and relative peak areas were calculated in the very fast application. All analytical results are compared with those obtained by conventional 0.25 mm ID column applications.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.